1. The RF Generator should be placed on a Stryker cart or on any sturdy table or platform. Refer to hospital
procedures or local codes for detailed information.
2. Provide at least four to six inches of space around the sides and top of the RF Generator for convection cooling.
It is normal for the top and rear panel to be warm when system is used continuously for extended periods of
time.
3. The RF Generator is shipped with an approved hospital grade mains power cord. Plug the RF Generator into a
grounded receptacle. DO NOT use extension cords or three-prong to two-prong adapters. The mains power cord
assembly should be checked periodically for damaged insulation or connectors.
CONNECTING SYCONNECTING SY
CONNECTING SYCONNECTING SY
CONNECTING SY
STEM STEM
STEM STEM
STEM
AA
AA
A
CCESSORIES:CCESSORIES:
CCESSORIES:CCESSORIES:
CCESSORIES:
NOTE: Refer to System Set-up Instructions located on pull-out tray.
GENERATOR
1. Plug power cord into back of Generator.
2. Plug power cord into AC mains.
FOOTSWITCH
1. Connect Footswitch cable to the Footswitch connector located on the front panel of the Generator.
2. When proper connection is achieved, the Footswitch indicator above the connector will illuminate.
SET UP AND INTERCONNECT SECTION
8
NOTE: Use only the the appropriate SERFAS accessories with the RF Generator. Refer to the Product Description
and Use section for product reference numbers.