7. RF COAG Output Level Indicators – The blue indicators corresponding to the relevant COAG buttons (HIGH,
MED, and LOW) will illuminate to indicate the level set.
8. Footswitch Connector – Connection for the footswitch. The Footswitch indicator above the connector will
illuminate upon proper connection to the RF Generator. The footswitch has four pedals that control CUT
Activation, COAG Activation, CUT Level Up, and CUT Level Down.
9. Footswitch Indicator – Indicator will illuminate upon proper connection of Footswitch to Generator.
10. Handpiece Cable Connector – Connection for the Handpiece Cable that connects to the Probe. The Handpiece
Cable facilitates radio-frequency energy delivery to the electrode tip, CUT activation, COAG activation and probe
family identification. The Probe indicator above the connector will illuminate upon proper connection of the
Handpiece Cable to the RF Generator and the Handpiece Cable to the Probe.
11. Probe Indicator – Indicator will illuminate upon proper connection of Handpiece Cable to Generator.
12. Error Indicator – A red indicator will flash to indicate an error.
REAR PANEL:
NOTE: Refer to Appendix B for rear panel labeling definitions.
1. Equipotential Ground Stud – Provides a means of securely linking the earth grounds of SERFAS to other
grounded equipment.
2. RS-232 Port – For manufacturing and test purposes only.
3. Power Entry Module – This module has a fuse drawer that contains two fuses.
4. Flash Memory Screw – For manufacturing and repair access only.
5. Speaker Volume Control – Controls the volume of the tones, beeps and voice feedback.
6. Voice/Tone Selector – A two-position switch, which enables or disables the digitized audio messages.
PRODUCT DESCRIPTION AND USE
1
2 5
6
3
4
5