Guidance and manufacturer's declaration - electromagnetic emissions
The TPS Console is intended for use in the electromagnetic environment specified below. The customer or the user of the TPS Console should
assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment - guidance
The TPS Console uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The TPS Console is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Frequency of transmission: 13.56MHz Subcarrier: 423.75kHz
Manchester coding
Type of Frequency/
Characteristics of the modulation: 10% ASK Effective radiated power: 50µW