Sun Nuclear 1027 - Updating USB to Serial Adapter Drivers; Driver Update Procedure

Sun Nuclear 1027
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
3 Click link below search box to display a list of available drivers, then click
Download link for your operating system.
4 In ‘File Download window,’ click Save, and save file to your hard drive. If
downloaded file is compressed (*.zip or *.rar format), use an unzip utility to
extract contents. Make note of where file is saved/extracted.
5 Disconnect and then reconnect USB end of adapter from computer. ‘Found
New Hardware Wizard’ is displayed.
6 Click Locate and install driver software (recommended). If Windows
requests permission to continue, click Continue. Windows may search for
drivers online and install them if it finds them. Otherwise, continue.
7 If ‘Insert the disc’ dialog box opens, click I don’t have the disc. Show me
other options.
8 If Windows can’t find the driver software, click Browse my computer for
driver software (advanced).
9 When prompted for driver location, click Browse and navigate to directory
where you saved/extracted driver files. If files are in subfolders for several
operating systems, choose correct subfolder for your operating system.
10 Click Next to install driver. When driver is installed, screen shows that the
software for the USB to Serial Adapter was successfully installed.
11 Click Close.
Troubleshooting Data Download Issues
The steps in this section are for users who have already followed the procedures
in Using USB to Serial Adapter on page 8 and have 1) already connected a USB to
Serial adapter, 2) installed drivers for the adapter following the cable
manufacturer’s instructions, and 3) are still unable to download data from Model
1027 to the computer.
Typically, the inability to download data from Model 1027 is caused by the Com
port number assigned to the adapter. The Radon Monitor software only
recognizes Com ports 1 through 4, so if the cable uses Com port 5 or higher the
software will not recognize the cable. Perform this procedure to change the Com
port assignment to 4 or lower.
1 Turn Model 1027 OFF and exit Radon Monitor software.
Note: Vista 64-bit drivers are not supported by Radon Monitor.
Note: To use this procedure you may need to be logged in to the
computer as an administrator.