EasyManuals Logo

Symmetricom 5071A Assembly And Service Manual

Symmetricom 5071A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #100 background imageLoading...
Page #100 background image
2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
84
A2 Hot-Wire Ionizer Diagnostic Tree
This functional subsection describes the A2 hot-wire ionizer diagnostic tree
for the 5071A Primary Frequency Standard. Performance of these tests is
indicated by the presence of either a self-test or fatal error that reports a hot-
wire ionizer fault.
3.1.1.1.0 CBT Hot-Wire Resistance Test
1 Power off the 5071A.
2 Unplug cable from A2J1 and A2J2.
3 Using a 4-wire Ohm Meter, measure the resistance across pins
1 and 6 on the CBT cable.
QUESTION: Is the resistance between 0.1 and 0.25 ?
If Yes: Go to, 3.1.1.1.1.
If No: Recheck results and Ohmmeter, then replace CBT.
3.1.1.1.1 CBT Hot-wire Resistance-to-Ground Test
1 Measure the resistance across pins 1 and 4 on the CBT cable.
QUESTION: Is the resistance greater than 200 k?
If Yes: Go to, 3.1.1.1.2.
If No: Recheck results and DMM, then replace CBT.
3.1.1.1.2 Hot-Wire Monitor Test (No Current)
1 Remove instrument power.
2 Remove A2 CBT Controller.
3 Install the long extender board (05071-60052) into A2 slot.
4 Turn on the 5071A.
5 Use SCPI command syst:print?.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION: Is HW Ionizer: between -0.5 and +0.5 V?
If Yes: Go to, 3.1.1.1.3.
If No: Remove power, replace A7, and retest.

Table of Contents

Other manuals for Symmetricom 5071A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom 5071A and is the answer not in the manual?

Symmetricom 5071A Specifications

General IconGeneral
BrandSymmetricom
Model5071A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals