2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
88
A2-Mass-Spectrometer Diagnostic Tree
This functional subsection describes the A2 mass-spectrometer diagnostic
tree for the 5071A Primary Frequency Standard. Performance of these tests
is indicated by the presence of either a
self-test or fatal error that reports a mass-spectrometer fault.
3.1.3.1.0 Mass-Spectrometer Resistance Test
1 Remove instrument power.
2 Unplug CBT cables from A2J1 and A2J2.
3 Use a DMM to measure the resistance across A2P4 CBT cable
pins 1 and 4.
QUESTION: Is the resistance greater than 200 k?
If Yes: Go to, 3.1.3.1.1.
If No: Recheck results and DMM, then replace CBT.
3.1.3.1.1 Mass-Spectrometer Test (Current Off)
1 Remove instrument power.
2 Replace A2 CBT Controller with long extender board (05071-60052).
(Do not insert A2 at this time.)
3 Reapply instrument power: instrument will go to Fatal Error.
4 Use SCPI command syst:print?.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION: Is Mass spec: between -0.5 and +0.5V?
If Yes: Go to, 3.1.3.1.2.
If No: Remove power, replace A7, and retest.
3.1.3.1.2 Mass-Spectrometer Monitor Test (+ Current On)
1 Place a 10 k resistor between pins 1 and 37 on connector XA2P2
(48-pin connector on extender card).
2 Use SCPI command syst:print?.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION: Is Mass spec: between 9.5 and 11V?
If Yes: Go to, 3.1.3.1.3.
If No: Remove power, replace A7, and retest.