EasyManuals Logo

Symmetricom 5071A Assembly And Service Manual

Symmetricom 5071A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #121 background imageLoading...
Page #121 background image
2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
105
3.5.4.2.1 A10 Port-2 Frequency Selection Error Test
1 Check SEL A on A10 with a DMM.
2 Verify that it is no greater than 2 volts.
QUESTION: Is the SEL A voltage 2 or less?
If Yes: Go to, 3.5.4.2.2.
If No: Remove power, replace A10, and retest.
3.5.4.2.2. A10 SEL-B Line Test
1 Remove instrument power.
2 Unsolder gray wire from SEL B terminal.
3 Reapply instrument power.
4 Check voltage on gray with DMM.
5 Verify that voltage is greater than 2 volts.
QUESTION: Is voltage greater than 2 volts?
If Yes: Remove power, replace A7, and retest.
If No: Remove power, replace A10, and retest.
3.5.5.0 A10 1 MHz Output Test
1 Connect the 1 MHz output connector on the rear panel to the 50 input of a
scope using coax. (If a 50 input is absent, use the high impedance input
with a 50 coaxial feedthrough terminator at the scope end of the coax).
2 Verify presence of a 1-MHz undistorted sine wave with at least
2.8 Vp-p amplitude.
QUESTION: Does the scope show a 1-MHz undistorted sine wave
with at least 2.8 Vp-p amplitude?
If Yes: Go to, 3.5.6.0
If No: Remove power, replace A10, and retest.

Table of Contents

Other manuals for Symmetricom 5071A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom 5071A and is the answer not in the manual?

Symmetricom 5071A Specifications

General IconGeneral
BrandSymmetricom
Model5071A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals