EasyManua.ls Logo

Symmetricom 5071A - Page 126

Symmetricom 5071A
202 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
110
3.6.5.0 A15J1 Power-Level Check
1 Connect A15J1 to a +2.5 Vdc power supply.
2 Check spectrum analyzer again and verify that amplitude has increased to at
least 0 dBm.
3 Disconnect power supply from A15J1 and reconnect internal cable to A15J1.
QUESTION: Did the amplitude on the spectrum analyzer increase
to at least 0 dBm?
If Yes: Go to, 3.6.6.0.
If No: Remove power, replace A15, and retest.
3.6.6.0 A15J2 Frequency-Range Test
1 Disconnect cable from A15J2.
2 Connect a 0 to +10 volt power supply to A15J2.
3 Vary voltage from 0 to +10V while watching signal on spectrum analyzer
connected to A15J5.
4 Verify that a frequency of 9192.6 MHz can be obtained with a voltage
between +2 and +8 Vdc.
QUESTION: Can frequency of 9192.6 MHz be obtained with a voltage
between +2 and +8 Vdc?
If Yes: Go to, A18 diagnostic tree in this section.
If No: Remove power, replace A15, and retest.

Table of Contents

Other manuals for Symmetricom 5071A

Related product manuals