EasyManuals Logo

Symmetricom 5071A Assembly And Service Manual

Symmetricom 5071A
202 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
53
1.3.3 CBT-Signal Fatal Errors
These errors occur when the instrument completes its lock-up procedure, and
the frequency servo loop cannot operate correctly. They usually indicate the
CBT output signal is unstable. This can be due to a malfunction in the CBT
itself, or more frequently, in the circuits associated with it. The following
procedure tests these alternatives.
1 Perform the Analog-Signal Chain tests (2.2.1.0) and repair as required.
2 Perform the RF-Chain tests (2.3.1.0) and repair as required.
3 Perform the A16 Electron-Multiplier Supply tests (3.7.2.1) and repair as
required.
4 Perform the RF-Level Control-Current Check, (1.3.8), and repair as required.
5 Apply power to the instrument: observe the warm up and subsequent
operation.
QUESTION: Did the instrument fail in the same manner?
If Yes: The CBT is possibly defective, contact Symmetricom for further
assistance.
If No: Problem may be resolved - retest.
1.3.4 CBT Shutdown Message
1 Read the CBT shutdown message preceding the error code.
QUESTION: What is the message?
If CBT shutdown:
thermistor open, Go to, A2 thermistor
(3.1.2.1.0) diagnostic tree.
If CBT shutdown:
thermistor shorted, Go to, A2 thermistor
(3.1.2.1.0) diagnostic tree.
If CBT shutdown:
ion pump overcurrent, Go to, Excess Ion-
pump current (1.3.2.1) test.
If CBT shutdown:
E-mult adjust failed – This indicates that although
larger e-mult movements were attempted the internal CBT parameters
could not be compensated sufficiently to allow continued operation.
Repair is likely necessary, however a re-start of the 5071A should first
be attempted as a possible remedy. If the failure re-occurs with the
same message, contact Symmetricom to arrange repair/replacement of
the CBT. If the error is not repeated but another error occurs, follow the
procedures for that error.

Table of Contents

Other manuals for Symmetricom 5071A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Symmetricom 5071A and is the answer not in the manual?

Symmetricom 5071A Specifications

General IconGeneral
BrandSymmetricom
Model5071A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals