SYNRAD® 32-1 Operator’s Manual Version 2.2
60
Troubleshooting
Maintenance/
Troubleshooting
Input/Output/Fault Conditions
LED and I/O States, 32-1 Series Lasers
Remote Key
PWM
Fault
Ready LED
Lase LED
DB-9 pins
6 to 7 3 to 4
1 9 8
DC Power Off X X
Laser enabled, no tickle or
PWM
C None H ON
Tickle Applied C Tickle H ON Tickle
Laser Firing C H ON
Over Temp Fault X X L
Under Voltage Fault X X L
Over Voltage Fault X X L
LASER CONDITION /
FAULT
Not Enabled
H ON
O X
L
Table 5-2 Input/Output/Fault Conditions for the 32-1 series lasers.