Operation
3
SYNRAD® 32-1 Operator’s Manual Version 2.2
27
Use information in this chapter to familiarize yourself with 32-1 controls and indicators. Refer-
ence the Quick Start Guide for the initial startup process.
This chapter contains the following information:
Controls and indicators – displays and describes exterior controls and indicators on 32-1
OEM lasers.
Initial start-up – Reference the Technical Reference chapter in this operation manual.
Caution
Remove the aperture seal before ring the laser. The self-adhe-
sive seal is installed to prevent dust from entering the laser hous-
ing during shipment and installation and must be removed before
operation. During laser operation, use a gas purge to keep dust and
vapor out of the beam path.
Applying PWM Command pulses directly to the laser without rst
sending tickle pulses for at least two seconds will cause unpredict-
able laser emission, degrade optical rise time, and may lead to RF
Driver failure.
Possible
Equipment
Damage