SYNRAD® 32-1 Operator’s Manual Version 2.2
56
Maintenance
Maintenance/
Troubleshooting
Cleaning guidelines
Before using any cleaning agents, read Safety Data Sheets (SDS) and observe all nec-
essary safety precautions.
Wear latex gloves or nger cots (powder-free) to prevent contamination of optical
surfaces by dirt and skin oils.
Never handle optics with tools; always use gloved hands or ngers.
Hold optics by the outer edge; never touch the coated surface.
Always place optics on a tissue or suitable equivalent material for protection; never
place optics on hard or rough surfaces.
It may be necessary to use a cotton ball or ued cotton swab instead of a lens wipe
to uniformly clean the entire surface of small-diameter mounted optics.
Required cleaning materials
The table below lists the type and grade of materials required to properly clean optical surfac-
es.
Table 5-1 Required cleaning materials.
Cleaning Material Requirements
Latex gloves or nger cots Powder free
Air bulb Clean air bulb
Ethyl or isopropyl alcohol Spectroscopic or reagent grade
Acetone Spectroscopic or reagent grade
Lens wipe (preferred) Optical (cleanroom) quality
Cotton balls or cotton swabs High-quality surgical cotton/high-quality paper-bodied