Connecting
7
Turn on the chiller and adjust the temperature setpoint to 18 °C to 22 °C. Regulate coolant flow
to 1.5 GPM at less than 60 PSI of pressure.
8
Closely examine all cooling connections and verify that there are no leaks.
Electrical connections
DC power supply
SYNRAD power supplies, models DC-1, DC-2, and DC-5, are sized to power 48-1, 48-2, and 48-5 lasers
respectively. The 48-1 laser requires a SYNRAD DC-1 or equivalent power supply capable of providing
30 VDC at 7 A maximum; the 48-2 laser requires a SYNRAD DC-2 power supply (30 VDC at 14 A
max.); and the 48-5 requires a SYNRAD DC-5 power supply (30 VDC at 28 A max.).
If substituting power supplies, use a well-regulated DC power supply in the range of 30 V to 32 V with no
more than 3 V overshoot under a 10–90% modulation load. The use of short leads and terminations
rated for the appropriate output current is recommended.
Note: SYNRAD lasers are tested to meet published specifications at an input voltage of 30 VDC.
AC input power requirements for the DC-1 power supply are 85–264 VAC, single phase (1Ø), 20 A
max, 47–440 Hz. Input requirements for both DC-2 and DC-5 supplies are 90–132 VAC (low range)
or 175–264 VAC (high range), single phase (1Ø), 13 A (DC-2) or 25 A (DC-5) max, 47–440 Hz.
DC-1 power supply
To connect the DC-1 power supply, refer to Figure 1-5 and perform the following steps:
Figure 1-5 DC-1 (Astec) power supply
V4
RET
V3
COM
COM
V2
V1
COM
OUTPUT
+
–
30 VDC
Output
AC
Input
Caution
possible
equipment
damage
Inlet cooling water temperature must always be maintained above
the dew point to prevent condensation and water damage to your
laser.