Performance Characteristics
Table 5-2
Environmental Characteristics
Characteristics
Description
Temperature
Operating
C to + 40°C (+ 32° F to + 104°F).
Nonoperating - 55°C to + 75°( - 67°F to
+ 167°F).
Altitude
Operating To 4500 m (15,000 ft.). Maximum
operating temperature decreases
1°C per 300 m (1,000 ft.) above
1,500 m (5,000 ft.).
Nonoperating To 15,250 m (50,000 ft.).
Relative Humidity
Operating
(+ 30°C to + 40°C)
95% +0%, -5 %
Nonoperating
(+ 30°C to + 60°C)
95% +0%, -5%
Vibration (Operating)
15 minutes along each of three
major axes at a total displacement
of 0.015 inch p-p (2.4 g at 55 Hz)
with frequency varied from 10 Hz to
55 Hz to 10 Hz in one-minute
sweeps. Hold for 10 minutes at
55 Hz in each of the three major
axes. All major resonances must be
above 55 Hz.
Shock (Operating and
Nonoperating)
30 g, half-sine, ll-ms duration, three
shocks per axis each direction, for
a total of 18 shocks.
Radiated and Conducted
Emission Reguirements
Meets VDE 0871 Class B and FCC
Regulations.
5-6
2225 Operators

Other manuals for Tektronix 2225

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2225 and is the answer not in the manual?

Tektronix 2225 Specifications

General IconGeneral
BrandTektronix
Model2225
CategoryTest Equipment
LanguageEnglish

Related product manuals