EasyManuals Logo
Home>Tektronix>Test Equipment>2236

Tektronix 2236 User Manual

Tektronix 2236
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Operators
Familiarization
--
2236
Operators
PROBE
COMPENSATION
O
Channel
1
and
Channel
2
CORRECT
(
FLAT
)
100
90
Misadjustment
of
probe
compensation
is
one
of
the
source
of
measurement
error
.
Most
attenuator
probes
are
equipped
with
a
compensation
adjustment
.
To
ensure
opti
mum
measurement
accuracy
,
always
compensate
the
oscil
loscope
probes
before
making
measurements
.
Probe
compensation
is
accomplished
as
follows
:
OVER
COM
PENSATED
(
OVERSHOOT
)
1.
Preset
instrument
controls
and
obtain
a
baseline
trace
.
10
1
n
Hei
1
UNDER
COM
PENSATED
(
ROLLOFF
)
2.
Connect
the
two
10X
probes
(
supplied
with
the
instru
ment
)
to
the
CH
1
and
CH
2
input
connectors
.
-
ji
(
4207-11
)
4205-17
3.
Set
both
VOLTS
/
DIV
switches
to
10
m
and
set
both
Input
Coupling
switches
to
AC
.
Fig
.
3-2
.
Probe
compensation
.
4.
Select
CH
2
VERTICAL
MODE
and
insert
the
tip
of
the
Channel
2
probe
into
the
PROBE
ADJUST
output
jack
.
1
Channel
1
AC
RMSV
5.
Using
the
approximately
1
-
kHz
PROBE
ADJUST
square
-
wave
signal
as
the
input
,
obtain
a
5
-
division
display
of
the
signal
.
Best
measurement
accuracy
above
1
kHz
with
a
10X
probe
can
be
accomplished
by
adjusting
the
probe
with
the
aid
of
a
precision
Ac
Calibration
System
.
Probe
compensa
tion
is
accomplished
as
follows
:
6.
Set
the
A
SEC
/
DIV
switch
to
display
several
cycles
of
the
PROBE
ADJUST
signal
.
use
the
Channel
2
POSITION
control
to
vertically
center
the
display
.
The
equipment
listed
in
Table
3-2
,
or
equivalent
equip
ment
,
is
required
to
compensate
the
10X
probe
for
ac
rms
volts
measurement
.
Table
3-2
Equipment
Required
for
10X
Probe
Compensation
7.
Check
the
waveform
presentation
for
overshoot
and
rolloff
(
see
Figure
3-2
)
.
If
necessary
,
adjust
the
probe
com
pensation
for
flat
tops
on
the
waveforms
.
Refer
to
the
in
structions
supplied
with
the
probe
for
details
of
compensation
adjustment
.
Description
П
Minimum
Specification
AC
Calibration
System
8.
Select
CH
1
VERTICAL
MODE
and
connect
the
Chan
nel
1
probe
tip
to
the
PROBE
ADJUST
output
jack
.
Ac
voltage
:
400
mV
to
4
V.
Voltage
accuracy
:
0.01
%
.
Frequency
:
10
kHz
.
Impedance
:
50
2.
Length
:
42
inch
.
Cable
9.
Use
the
Channel
1
POSITION
to
vertically
center
the
display
and
repeat
step
7
for
the
Channel
2
probe
.
Connectors
:
BNC
.
Adapter
Connectors
:
BNC
-
male
-
to
-
miniature
probe
tip
.
NOTE
П
Channel
1
ac
rms
volts
measurement
accuracy
above
1kHz
depends
directly
upon
the
precision
with
which
the
preceding
adjustment
is
performed
.
An
overshoot
of
0.2
division
(
or
4
%
of
a
5
-
division
display
)
will
add
approximately
4
%
to
the
displayed
reading
.
Compen
sation
related
errors
can
be
reduced
to
1
/
2
%
or
less
by
carefully
adjusting
the
probe
compensation
to
the
PROBE
ADJUST
square
-
wave
signal
.
1.
Preset
instrument
controls
and
obtain
a
baseline
trace
,
set
CH
1
VOLTS
/
DIV
switch
to
50
mV
,
and
Channel
1
Input
Coupling
switch
to
AC
RMSV
.
C
2.
Connect
the
AC
Calibration
System
output
via
a
50-1
cable
to
CH
1
OR
X
&
DMM
input
connector
.
REV
JUL
1984
3-5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2236 and is the answer not in the manual?

Tektronix 2236 Specifications

General IconGeneral
BrandTektronix
Model2236
CategoryTest Equipment
LanguageEnglish

Related product manuals