Performance Tests (MSO70000/C Series, DSA/DPO70000B/C Series, and DPO7000 Series)
Minimum requir
ements
Example Purpose
21. Cable, Coaxia
l(three
required)
50 Ω,20in,mal
e-to-male
SMA connectors
Tektronix part number
174-1427-00
Used to test delta time
measurement a
ccuracy,
probe compensation out,
trigger sensitivity
22. Adapter
SMA "T", male to 2 SMA
female
Tektronix part number
015-1016-00 (18 GHz)
Used to test delta time
measurement accuracy
23. Adapter
SMA female t
oBNCmale
Tektronix part number
015-0572-0
0(4GHz)
Used to test delta time
measuremen
t accuracy and
trigger sensitivity
24. Adapter
BNC male to female elbow
Tektronix p
art number
103-0031-00
Used to test
delta time
measurement accuracy
25. Terminati
on
Short circuit, SMA connector,
female
Tektronix
part number
015-1021-00 (18 GHz)
Used to tes
t delta time
measurement accuracy
26. Attenuator, 2X
Ratio: 2X; impedance 50 Ω;
connecto
rs: female BNC
input, male BNC output
Tektronix part number
011-0069-02 (2 GHz)
Used to test delta time
measurement accuracy,
pulse tri
gger accuracy, time
qualified trigger, trigger
sensitivity, and channel
isolati
on
27. Digital Multimeter
Ohms: <6
0Ohms
Keithley 2700
Checkin
g input impedance
28. Cable,
coaxial
50 Ω,39
.37in(1.0m),
male-to-male SMA connectors
50 Ω,60in(1.5m),
male-
to-male S MA connectors
Tektronix part number
174-13
41-00
Tektronix part number
174-1428-00
Checki
ng analog bandwidth
and delta time measurement
accuracy. Checking ≥ 4GHz
model
s serial trigger baud
rate limits
29. Mouse
or keyboard
Tektr
onix part numbers:
119-6298-xx (mouse)
119-6297-xx (keyboard)
Used t
o input test selections
30. Probe tip adapter
BNC female to square pin
adap
ter, 50 Ω
Tektronix part number
067–1734–00
Checking digital channels
31. Logic probe P6717 or P6780 Logic Probe Tektronix part P6717 or
P67
80
Che
cking digital channels
1
Fluke 9500B/1100, 9500B/3200, or 9500B/2200 and an output head (9510 or 9530) appropriate for the bandwidth of the instrument being tested.
2
For Delta Time Measurement Accuracy, use a Fluke 9500B or a pulse generator with the specified rise time. (See Table 3-10 on page 3-78.)
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series 3–21