EasyManuals Logo

Tektronix DPO70000 Series Technical Reference

Tektronix DPO70000 Series
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #245 background imageLoading...
Page #245 background image
Performance Tests (MSO/DPO5000 Series)
8. Record the nega
tive-measured and positive-measured mean readings in the
worksheet as follows:
a. Set the DC Volt
ageSourcetoV
negative
.
b. Push the front-panel Wave Inspector Measure button.
c. In the Measurements Setup window, go to the Measurements box. Click
the More button and then select Mean.
d. View the Mean measurement value in the display.
e. In the Measurement Setup window, select Statistics.
f. Click Reset.
g. Enter the mean reading in the worksheet as V
negative-measured
.(SeeTable4-2
on page 4-32.)
h. Set the DC Voltage Source to V
positive
.
i. Click R
eset again.
j. Enter the mean reading in the worksheet as V
positive-measured
.
MSO70000/C, DSA70000B/C, DPO7000B/C, DPO7000, MSO5000, DPO5000 Series 4–31

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals