EasyManuals Logo
Home>Tektronix>Test Equipment>DPO7054

Tektronix DPO7054 Quick Start User Manual

Tektronix DPO7054
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Acquisition
3. When the status
changes to Temp,
click Calibrate to start the calibration.
Calibration m ay take 10 to 15 minutes.
4. If the i nstrument does not pass,
recalibrate the instrument, or have the
instrumen
t serviced by qualied service
personnel.
NOTE. To always show the SPC status icon
or display
a warning when SPC has not
been run in more than a month, click the
corresponding option button.
5. If the SPC needed icon is red, perform a
signal path compensation.
24 DPO7000 Series and DPO /DS A70000/B Series Quick Start User Manual

Table of Contents

Other manuals for Tektronix DPO7054

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7054 and is the answer not in the manual?

Tektronix DPO7054 Specifications

General IconGeneral
BrandTektronix
ModelDPO7054
CategoryTest Equipment
LanguageEnglish

Related product manuals