Acquisition
Acquisition
This section contains concepts of and procedures for using the acquisition system.
Signal Path Compensation
Use this procedure if the temperature has changed more than 5 °C (9 °F) since the last signal path c ompensation.
Perform the s
ignal path compensation once a week. Failure to do so may result in the instrument not meeting warranted
performance levels.
1. Prerequisites: instrument powered on
for 20 minutes, and all input signals
removed.
2. Select In stru men t Calibration.
DPO7000 Series and DPO/DSA70000/B Series Quick Start User Manual 23