Commands Listed in Alphabetical Order
HORizontal:M
ATH<x>:MAGnify<x>:RECordlength? (Query Only)
This query only command r eturns the record length of the specified Mag timebase
for the specified math waveform. The value of the record length can be 20, 50,
100, 250, 500
, 1000, 2000, and 4000 samples. It results from normalizing the
sources of a math waveform.
Group
Horizontal
Syntax
HORizontal:MATH<x>:MAGnif y<x>:RECordlength?
Related Commands
HORizontal:MATH<x>:MAGnify<x>:SCAle?, HORizontal:MATH<x>:
MAGnify<x>:RESolution?
Examples
HORIZONTAL:MATH5:MAG2:REC ORDLENGTH? might return
HORIZONTAL:MATH5:MAGNIFY2 :RECORDLENGTH 1000, indicating
thattherecordlengthoftheMag2timebase is set to 1000 samples for the Math
5 waveform.
HORizontal:MATH<x>:MAGnify<x>:RESolution? (Query Only)
This query only command returns the resolution per sample of the specified Mag
timebase for the specified math waveform. The math waveform is specified by x,
which can be 1 through 8; the Mag timebase is specified by x, which can be 1 or
2. Resolution is displayed in horizontal units, typically time.
Group
Horizontal
Syntax
HORizontal:MATH<x>:MAGnif y<x>:RESolution?
Related Commands
HORizontal:MATH<x>:MAGnify<x>:SCAle?, HORizontal:MATH<x>:
MAGnify<x>:RECordlength?
Examples
HORIZONTAL:MATH1:MAG1:RES OLUTION? might return
HORIZONTAL:MATH1:MAGNIFY1 :RESOLUTION 1.00000000000 E–009,
indicating that the resolution of the Mag1 timebase for the Math1 waveform is
1ns.