EasyManuals Logo

Tektronix DSA8200 Series User Manual

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #242 background imageLoading...
Page #242 background image
Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>:MEAN? (Query Only)
This query only command returns the mean value accumulated for the
measurement slot specied by x, which c an be 1 through 8, since the last statistical
reset.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEAN?
Related Commands
MEASUrement:MEAS<x>:ALL?
Returns
NR3
Examples
MEASUREMENT:MEAS1:MEAN? might return ME ASUREMENT:MEAS1:MEAN
2.85516357276E-003
, indicating that the mean value for Meas 1 is 2.855 mV.
MEASUrement:MEAS<x>:MINimum? (Query Only)
This query only command returns the minimum value found for the measurement
slot specied by x, which can be 1 through 8, since the last statistical reset.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MINim um?
Related Commands
MEASUrement:MEAS<x>:ALL?
Returns
NR3
Examples
MEASUREMENT:MEAS1:MINIMUM ? might return
MEASUREMENT:MEAS1:MINIMUM 1.92260747428E- 003, indicating that the
minimum value for Meas 1 is 1.923 mV.
MEASUrement:MEAS<x>:NOISe
This command sets or queries whether the noise measurement is made on the high
or low level of the waveform. Sending this command is equivalent to selecting
2-228 DSA/CSA/TDS8X00/B Series Programmer Manual