EasyManuals Logo

Tektronix DSA8200 Series User Manual

Tektronix DSA8200 Series
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #238 background imageLoading...
Page #238 background image
Commands Listed in Alphabetical Order
MEASUrement:
LISTValue? (Query Only)
This query only command returns the values of the measurements in the list
created with the MEASUrement:LIST command.
Group
Measurement
Syntax
MEASUrement:LISTValue?
Related Commands
MEASUrement:LIST
Examples
MEASUREMEN
T:LISTVALUE?
might return MEASUREMENT:LISTVALU E
1.52587894784E-004,2.4949 9996255E-010
, which are the mea surement
values for the measurement slots dened as part of the list.
MEASUrem
ent:MEAS<x>:ALL? (Query Only)
This query only command returns all the statistics for the measurement
specied by x, which can be 1 through 8. Before you send this command,
send the
MEASUrement:STATIstics:ENABle ON command to ensure accurate
measurement statistics. The following measurement values are returned:
Maximu
m
Minimum
Mean
Standard deviation
Measurement value
Group
Measurement
Syntax
MEASUrement:MEAS<x>:ALL?
Related Commands
MEASUrement:MEAS<x>:MAXimum?, MEASUrement:MEAS<x>:
MINimum?, MEASUrement:MEAS<x>:MEAN?, MEASUrement:MEAS<x>:
STDdev?, MEASUrement:MEAS<x>:VALue?, MEASUrement:STATIstics:
ENABle
Re
turns
NR3
2-224 DSA/CSA/TDS8X00/B Series Programmer Manual

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals