Commands Listed in Alphabetical Order
NOTE. The SOURC
E<x> SOURCE:<<x> argument species the
source, which can be source 1 or source 2. To set the actual waveform
(channel, reference, or math) that is assigned to source, you must use the
MEASUrement:MEAS<x>:SOURCE<x> WFM command. Two sources are
provided because some measurements (gain, phase) require two sources.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce <x>:HILow:
TRACk:LOW:ENABle { ON
|OFF|NR1}
MEASUrement:MEAS<x>:SOUrce <x>:HILow:TRACk:LOW:ENABle?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:HILow:METHod, MEASUrement:
MEAS<x>:SOUrce<x>:HILow:TRACk:HIGH:ENABle
Arguments
ON enables tracking of the low value for the specied measurement on the
specied source waveform.
OFF disables tracking of the low value for the specied measurement on the
specied source waveform.
NR1 set to 0 disables tracking of the low value for the specied measurement
on the specied source wave form; any other value enables tracking of the low
value for the specied measurement on the specied source waveform.
Examples
MEASUREMENT:MEAS2:SOURCE1: HILOW:TRACK:LOW:ENABLE ON enables
tracking of the low value used to take Measurement 2 on the Source 1 waveform.
MEASUREMENT:MEAS3:SOURCE2: HILOW:TRACK:LOW:ENABLE? might return
MEASUrement:MEAS3:SOURCE2:HILow:TRACK:LOW:ENABle 0, indicating
that the tracking of the low value for Measurement 3 on the Source 2 waveform is
disabled.
NOTE. A query always returns a 0 or 1.
MEASUrement:MEAS<x>:SOUrce<x>:HILow:TRACk:LOW:VALue
This command sets or queries the low value used to calculate a specied
measurement on a specied source waveform. The measurement slot i s specied
by x, which c an be 1 through 8; the source is specied by x, which can be 1 or
2. Sending this command is equivalent to selecting a measurement slot (such
as Meas 3), selecting a source (such as Source1), setting tracking Low to off,
and then entering an absolute value in the Low control box in the in the Hi/Low
DSA/CSA/TDS8X00/B Series Programmer Manual 2-247

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA8200 Series and is the answer not in the manual?

Tektronix DSA8200 Series Specifications

General IconGeneral
BrandTektronix
ModelDSA8200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals