Commands Listed in Alphabetical Order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:STATE {OFF | ON | <NR1>}
MEASUrement:MEAS<x>:STATE ?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>:WFM, MEASUrement:MEAS<x>:TYPe
Arguments
OFF disables the calculation and display of the measurement denedinthe
specied measurement slot.
ON enables the calculation and display of the measurement dened in the
specied measurement slot.
NR1 set to 0 disables the calculation and display of the specied mea surement
slot; any other value enables the calculation and display of the specied
measurement slot.
Examples
MEASUREMENT:MEAS1:STATE ON displays the measurement dened in
measurement slot 1.
MEASUREMENT:MEAS4:STATE? might return MEASUREMENT:MEAS4:STA TE 1,
indicating that the measurement dened for measurement slot 4 is displayed.
NOTE. A query always returns a 0 or 1.
MEASUrement:MEAS<x>:STATIstics:CLEar (No Query Form)
This command (no query form) clears and restarts the calculation of the statist
ics
for the measurement slot specied by x, which can be 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:STATI stics:CLEar
Related Commands
MEASUrement:STATIstics:ENABle
Examples
MEASUREMENT:MEAS3:STATIST ICS:CLEAR clears and restarts calculating the
statistics for Measurement 3.
2-252 DSA/CSA/TDS8X00/B Series Programmer Manual