Commands Listed in Alphabetical Order
Syntax
TRIGger:HOLDo
ff <NR3>
TRIGger:HOLDoff?
Related Commands
HORizontal:MAIn:SCAle, HORizontal:MAIn:POSition, TRIGger:INTRate,
CH<x>:DESkew
Arguments
NR3 is the value of the holdoff time.
Examples
TRIGGER:HOLDOFF 5.0E-6 sets the requested trigger holdoff to 5 µs.
TRIGGER:HOLDOFF? might return TRIGGER:HOLDOFF 6.00000000000E-006,
indicating that the actual holdoff is set to 6 µs.
TRIGger:INTRate
This command sets or queries the internal clock rate. The command sets the same
parameter as the TDR:INTRate command. The internal clock in the instrument
generates a continuous signal with selectable repetition rates of 200 kHz (default),
100 kHz, 50 kHz, and 25 kHz. Sending this command is equivalent to selecting
a value for Internal Clock in the Trigger or TDR Setup dialog box. Use th e
TRIGger:SOUrce command to select the internal clock. This interna l clock rate
can be used to synchronize the generation of TDR step pulses with the subsequent
acquisition.
Group
Trigger
Syntax
TRIGger:INTRate <NR3>
TRIGger:INTRate?
Related Commands
TDR:INTRate
Arguments
NR3 sets the internal clock rate to one of the following: 25 kHz, 50 kHz, 100 kHz,
or 200 kHz. You can input any NR3 value; the instrument will set the internal
clock rate to the closest available setting.
Examples
TRIGGER:INTRATE 100E3 sets the internal clock rate to 100 kHz.
TRIGGER:INTRATE? might return TRIGGER :INTRATE
200.00000000000E+003
, indicating that the internal clock rate is
set to the default, 200 kHz.
2-332 DSA/CSA/TDS8X00/B Series Programmer Manual