7½ Digit Multimeter User's Manual Section 4: Making basic front-
DMM7510-900-01 Rev. C / September 2019 4-3
To change the measure settings:
1. Press the MENU key.
2. Under Measure, select Settings.
3. Select Display Digits.
4. Select 3.5 Digits.
5. Press the HOME key. The measurement now shows 3½ digits.
To make a single measurement:
1. Press the front-panel TRIGGER key for a few seconds.
2. Select Manual Trigger Mode.
3. Press the TRIGGER key to initiate a single reading using the selected measure function.
Figure 23: Basic measurement test results
View measurement data
You can view data from the reading buffers through the front panel using the Reading Table. The
Reading Table displays the following information:
• Index: The sequential number of the reading.
• Time: The date and time of the reading.
• Reading: The data that was measured.
• Extra: Only displayed for buffers that are set to Full. The extra value that is stored with a reading,
such as the ratio component of a DCV ratio measurement.