Section
10: Capturing and analyzing waveforms Model DMM7510 7½ Digit Multimeter
User's Manual
10-2 DMM7510-900-01 Rev. C / September 2019
Device connections
This application example consists of four tests with different connections. An overview of the
connections is shown in the figure below.
You can use either the front-pane or rear-panel connections for these tests. Check the front panel
TERMINALS switch to verify that the instrument is set to read the correct set of terminals (F is lit for
front terminals; R is lit for rear terminals). You must use either the front terminals or rear terminals.
You cannot mix connections.
Figure 36: Buck converter evaluation board connections overview