Model DMM7510
7½ Digit Multimeter User's Manual Section 8:
Grading and binning resistors
DMM7510-900-01 Rev. C / September 2019 8-7
-- Configure digital I/O line 5 as a trigger input used to detect
-- the start-of-test trigger from the component handler.
digio.line[5].mode = digio.MODE_TRIGGER_IN
-- Set trigger detector to detect falling edge.
trigger.digin[5].edge = trigger.EDGE_FALLING
-- Configure digital I/O line 6 as a trigger output used to send
-- an end-of-test trigger to the component handler.
digio.line[6].mode = digio.MODE_TRIGGER_OUT
-- Output a falling edge trigger.
trigger.digout[6].logic = trigger.LOGIC_NEGATIVE
-- Set width of output trigger pulse to 10 us.
trigger.digout[6].pulsewidth = 10E-6
-- Trigger pulse is output when the notify block generates an event.
trigger.digout[6].stimulus = trigger.EVENT_NOTIFY2
-- Load the GradeBinning trigger-model template.
trigger.model.load("GradeBinning", 100, 5, .1, .1, 120, 80, 15, 4, 110, 90, 1, 105, 95,
2, 101, 99, 3, bufferVar)
-- Initiate trigger model and wait until finished.
trigger.model.initiate()
waitcomplete()
-- Display on front-panel USER swipe screen after binning test completes.
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Test Completed")

Table of Contents

Other manuals for Tektronix Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley DMM7510 and is the answer not in the manual?

Tektronix Keithley DMM7510 Specifications

General IconGeneral
BrandTektronix
ModelKeithley DMM7510
CategoryMultimeter
LanguageEnglish

Related product manuals