DMM7510-900-01 Rev. C / September 2019 9-9
node[2].trigger.model.setblock(14, node[2].trigger.BLOCK_NOTIFY,
node[2].trigger.EVENT_NOTIFY1)
node[2].trigger.model.setblock(15, node[2].trigger.BLOCK_BRANCH_COUNTER,
muxChannelsCount[3], 12)
node[2].trigger.model.setblock(16, node[2].trigger.BLOCK_CONFIG_NEXT,
"myScanConfigList")
node[2].trigger.model.setblock(17, node[2].trigger.BLOCK_WAIT,
node[2].trigger.EVENT_TSPLINK1)
node[2].trigger.model.setblock(18, node[2].trigger.BLOCK_MEASURE_DIGITIZE)
node[2].trigger.model.setblock(19, node[2].trigger.BLOCK_NOTIFY,
node[2].trigger.EVENT_NOTIFY1)
node[2].trigger.model.setblock(20, node[2].trigger.BLOCK_BRANCH_COUNTER,
muxChannelsCount[4], 17)
-- Initiate trigger models on both 3706A-S and DMM7510 and wait until finished.
node[2].trigger.model.initiate()
-- To guarantee that the DMM7510 trigger model is at WAIT block (block 2).
delay(2)
node[1].scan.background()
waitcomplete()
-- Open all channels on slots 1 and 4.
node[1].channel.open("slot1, slot4")