Series 2600B System SourceMeter® instrument User's Manual Section 3: Instrument description
2600BS-900-01 Rev. A / August 2021 3-43
Event log descriptions
Column title
Description
Example
Sequence
Each instrument maintains independent
sequence counters:
One for each combination of UDP multicast
network interface and UDP multicast
destination port
One for each TCP connection
Domain
Displays the LXI domain number; the default value
is 0 (zero)
0
Flags
Contain data about the LXI trigger packet;
values are:
1 - Error
2 - Retransmission
4 - Hardware
8 - Acknowledgments
16 - Stateless bit
16
Data
The values for this are always 0 (zero)
Accessing the event log from the command interface
You can access the event log from any remote command interface. The event log must be enabled
before LXI trigger events can be viewed. To enable the event log, send:
eventlog.enable = 1
To view the event log from a remote interface, send:
print(eventlog.all())
This command outputs one or more strings similar to the following:
14:14:02.000 17 Jun 2019, LAN0, 10.80.64.191, LXI, 0, 1560780842, not available, 0,
0x10,0x00

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix KEITHLEY SourceMeter 2600B Series and is the answer not in the manual?

Tektronix KEITHLEY SourceMeter 2600B Series Specifications

General IconGeneral
BrandTektronix
ModelKEITHLEY SourceMeter 2600B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals