Performance Ver
ication
6. Check from 5 MHz to BW (MDO 3SA not installed):
a. Set the stop frequency to the maximum bandwidth.
b. Set the start frequency to 5 M Hz.
c. Set Manual Marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Set the center frequency as follows: Push the RToCenterside-bezel button.
e. Set the span to 10 MHz as follows: Push the side-bezel Span button. Set the Span to 10 MHz.
f. Record the highest noise value (in dBm/Hz) in the test record and compare it to the instrument specication.
7. Check from 5 MHz to 2 GHz (MDO3SA installed).
a. Set the stop frequency to 2 G Hz.
b. Set the start frequency to 5 M Hz.
c. Set Manual Marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Set the center frequency as follows: Push the RToCenterside-bezel button.
e. Set the span to 10 MHz as follows: Push the side-bezel Span button. Set the Span to 10 MHz.
f. Record the highest noise value (in dBm/Hz) in the test record and compare it to the instrument specication.
8. Check from 2 GHz to 3 GHz (MDO3SA installed).
a. Set the stop frequency to 3 G Hz.
b. Set the start frequency to 2 GHz.
c. Set Manual Marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Set the center frequency as follows: Push the RToCenterside-bezel button.
e. Set the span to 10 MHz as follows: Push the side-bezel Span button. Set the Span to 10 MHz.
f. Record the highest noise value (in dBm/Hz) in the test record and compare it to the instrument specication.
This completes the procedure.