Performance Ver
ification
c. Set Manual Marker (a) at the frequency of the highest noise, ignoring any spurs.
d. Set the center frequency as follows: Push the RToCenterside-bezel button.
e. Set the Span to 10 MHz as follows.
Push the side-bezel Span button.
Set the Span to 10 MHz.
f. Record the highest noise value (in dBm/Hz) in the test record and compare it to the instrument specification.
This completes the procedure.
108 MDO3000 Series Specifications and Performance Verification