Specifications
Table 1: Analog channel input and vertical specifications (cont.)
Characteristic Description
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Vertical Scale Setting
Instrument
Bandwidth
10 mV/div to
1V/div
5 mV/div to
9.98 mV/div
2 mV/div to
4.98 mV/div
1 mV/div to
1.99 mV/div
1.00 GHz DC to 1.0 GHz DC to 500 MHz DC to 350 MHz DC to 150 MHz
500 MHz
DC to 500 MHz DC to 350 MHz DC to 150 MHz
350 MHz
DC to 350 MHz DC to 350 MHz DC to 150 MHz
200 MHz
DC to 200 MHz DC to 150 MHz
Analog bandwidth,
50 input ter
mination
(See page 76.)
100 MHz
DC to 100 MHz
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Vertical Scale Setting
Instrument
Bandwidth
10 mV/div to
1V/div
5 mV/div to
9.98 mV/div
2 mV/div to
4.98 mV/div
1 mV/div to
1.99 mV/div
500 MHz,
350 MHz,
and 200 MHz
DC to 200 MHz DC to 140 MHz DC to 100 MHz
Analog bandwidth, 75
input termination, typical
100 MHz
DC to 100 MHz
The limits stated below are for ambient temperature of 30 °C and the bandwidth selection set to
FULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Vertical Scale Setting
Instrument
Bandwidth
2mV/divto
10 V/div
1 mV/div to
1.99 mV/div
1GHz,
500 MHz, or
350 MHz
DC to 350 MHz DC to 150 MHz
200 MHz
DC to 200 MHz DC to 150 MHz
Analog bandwidth, 1 M
input termination. The
Analog Bandwidth when
the instrument is DC-1M
coupled, typical
100 MHz
DC to 100 MHz
4 MDO3000 S eries Specifications and Performance Verification