Commands Listed in Alphabetical Order
Returns
This query will
return one of the following:
INIT indicates the RF portion of the instrument has not been calibrated.
PASS indicates the RF internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
FAIL indicates the RF internal calibration did not complete successfully.
Examples
CAL:INTERN:STAT:RF? might return INIT indicating that the RF portion of the
instrument has not been internally calibrated.
CALibrate:INTERNal:STATus:SCOPE? (Query Only)
This query returns the status of the last SPC run for the oscilloscope portion of
the instrument (doesn't include the RF portion). To query the status of the RF
portions, use CALibrate:INTERNal:STATus:RF?
This query is synonymous to CALibrate:RESults:SPC:SCOPE?
Group
Calibration and Diagnostic
Syntax
CALibrate:INTERNal:STATus:SCOPE?
Returns
Th
is query will return one of the following:
INIT indicates the oscilloscope portion of the instrument has not been calibrated.
PASS indicates the oscilloscope internal calibration completed successfully.
RUNNING indicates the oscilloscope internal calibration is currently running.
FAIL indicates the oscilloscope internal calibration did not complete successfully.
Examples
CAL:INTERN:STAT:SCOPE? might return I NIT indicating that the oscilloscope
portion of the instrument has not been calibrated.
CALibrate:RESults? (Query Only)
Returns the status of internal and factory calibrations, without performing any
calibration operations. The results returned do not include the calibration status
of attached probes. The
CALibrate:RESults? query is intended to support
GO/NoGO testing of the oscilloscope calibration readiness: all returned results
should indicate PASS status if the oscilloscope is "fit for duty". It is quite
2-214 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual