Commands Listed in Alphabetical Order
Returns
<NR1> = 1 if the S
PI bus feature is present.
<NR1> = 0 if the SPI bus feature is not present.
CONFIGuration:BUSWAVEFORMS:USB? (Query Only)
This query returns a boolean value to indicate whether the USB bus triggering
and analysis feature is present.
Conditions
This feature requires installation of a DPO4USB application module, or
MDO3USB for MDO3000 models.
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:USB?
Related Commands
(See page 2-15, Bus Command Group.)
(See page 2-76, Trigger Command Group.)
(See page 2-61, Search Command Group.)
Returns
<NR1> = 1 if the USB bus feature is present.
<NR1> = 0 if the USB bus feature is not present.
CONFIGuration:BUSWAVEFORMS:USB:HS? (Query Only)
This query returns a boolean value to indicate whether the high-speed USB bus
triggering and analysis feature is present.
Depending upon the bandwidth of the instrument, USB bus triggering and
analysis features may be limited to USB low-speed or full-speed. If the instrument
bandwidth is sufficient, USB high-speed (HS) triggering and analysis is supported
as long as the DPO4USB or MDO3USB application module is installed.
Group
Configuration
Syntax
CONFIGuration:BUSWAVEFORMS:USB:HS?
Related Commands
(See page 2-15, Bus Command Group.)
2-250 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual