Commands Listed in Alphabetical Order
Arguments
ZERo, OFF,and0
specify that the BRS bit must have value 0. If queried, the
command will always return 0 if set with these arguments.
ONE, ON,and1 s
pecify that the BRS bit must have value 1. If queried, the
command will always return 1 if set with these arguments.
NOCARE, X,a
nd
x specify that the BRS bit may be either 1 or 0. If queried, the
command will always return X if set with these arguments.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FD:ESIBIT
This command specifies the binary data value to be used to search on CAN FD
ESI bits. This only applies if the search condition has been set to FDESIBIT
(using
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:CONDition).
SEARCH:<x> is the search number, which is always 1, and B<x> is the bus
number (1-3), except for the MDO3000 which only has two buses.
Conditions
This command requires a DPO4AUTO or DPO4AUTOMAX application module
on MDO4000C models and a MDO3AUTO application module on MDO3000
models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:FD:
ESIBIT{ZERo|OFF|0|ONE|ON|1|NOCARE|X|x}
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CA N:FD:ESIBIT?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:CONDition
Arguments
ZERo, OFF,and0 specify that the BRS bit must have value 0. If queried, the
command will always return 0 if set with these arguments.
ONE, ON,and1 specify that the BRS bit must have value 1. If queried, the
command will always return 1 if set with these arguments.
NOCARE, X,andx specify that the BRS bit may be either 1 or 0. If queried, the
command will always return X if set with these arguments.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:CAN:FRAMEtype
This command sets the frame type (data, remote, e
rror or overload) to be used
to search on CAN bus data. This only applies if the search condition has been
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-629

Table of Contents

Other manuals for Tektronix MDO3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO3000 Series and is the answer not in the manual?

Tektronix MDO3000 Series Specifications

General IconGeneral
BrandTektronix
ModelMDO3000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals