Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Search
Syntax
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:ET HERnet:IPHeader:
PROTOcol:VALue <QString>
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:ET HERnet:IPHeader:
PROTOcol:VALue?
Related Commands
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:ETHERnet:CONDition
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:ETHERnet:IPHeader:
SOUrceaddr:VALue
SEARCH:SEARCH<x>:TRIG:A:BUS:B<x>:ETHER:IPH:DEST:VAL
Arguments
QString is a quoted string of up to 8 characters where the allowable characters
are 0, 1, and X. The bits specified in the quoted string replace the least significant
bits, leaving any unspecified upper bits unchanged.
Examples
SEARCH:SEARCH1:TRIGGER:A:BUS:B1:ETHERN ET:IPHEADER:
PROTOCOL:VALUE
“01010010” would set the value to be used in the search to
01010010.
SEARCH:SEARCH1:TRIGger:A:BUS:B1:ETHERn et:IPHeader:
PROTOcol:VALue?
might return “XXXXXXXX”.
SEARCH:SEARCH<x>:TRIGger:A:BUS:B<x>:ETHERnet:IPHeader:SOUrceaddr:VALue
When the Ethernet search condition is set to IPHeader, this command specifies
the value of the 32–bit source address that will be used in the search (along with
the protocol and destination address). TheIPsourceaddressisastandardIP
address such as 192.168.0.1. The default is all X’s (don’t care).
SEARCH<x> is the
search number, which is always 1, and
B<x> is the bus number 1-4. 1-3 for the
MDO4000C and 1-2 for the MDO3000.
Conditions
This command requires a DPO4ENET application module for DPO/MSO4000B
and MDO4000/B/C models.
Group
Search
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-637