Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:DATa:HIVALu e <QString>
TRIGger:A:BUS:B<x>:FLEXray:DATa:HIVALu e?
Arguments
<QString> is a quoted string that is the binary data high value.
Examples
TRIGGER:A:BUS:B1:FLEXRAY:DATA:HIVALUE
"11001101XXXXXXXXXXXXXXXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXXXXX
XXXXXXX1"
sets the binary data string high value to
"11001101XXXXXXXXXXXXXXXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXXXXX
XXXXXXX1"
.
TRIGGER:A:BUS:B1:FLEXRAY:DATA:HIVALUE? might
return
TRIGGER:A:BUS:B1:FLEXRAY:DATA:HIVALUE
"XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXXXXX
XXXXXXXX"
indicating the binary data string high value is don't care.
TRIGger:A:BUS:B<x>:FLEXray:DATa:OFFSet
This command specifies the offset of the data string, in bytes, when triggering
on the FlexRay bus data field. The trigger condition needs to be set to ID or
IDANDDATA (using TRIGger:A:BUS:B<x>:FLEXray:CONDition).
B<x>
Conditions
This command requires a DPO4AUTOMAX application module for
DPO/MSO4000B and MDO4000/B/C models or MDO3FLEX application module
for MDO3000 models.
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:FLEXray:DATa:OFFSet <NR1>
TRIGger:A:BUS:B<x>:FLEXray:DATa:OFFSet ?
Arguments
<NR1> is the offset of the data string in bytes. A byte offset of -1 signifies “don't
care”,andnobyteoffsetisused. Theinstrument will trigger on or match any
byte value that fits.
2-812 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual