Commands Listed in Alphabetical Order
Group
Trigger
Syntax
TRIGger:A:BUS:B<x>:MIL1553B:CONDition
{SYNC|COMMAND|STATus|DATA|TIMe|ERRor}
TRIGger:A:BUS:B<x>:MIL1553B:CONDition?
Related Commands
Most of the other TRIGger:A:BUS:B <x>:MIL1553B commands are impacted
by the setting of this command.
BUS:B<x>:MIL1553B:POLarity
BUS:B<x>:MIL1553B:SOUrce
Arguments
SYNC refers to the 3–bit sync pulse that precedes each word.
COMMAND is one of 3 16–bit word types. It specifies the function that a remote
terminal is to perform.
STATus is one of 3 16–bit word types. Remote terminals respond to valid message
transmissions via status words.
DATA is one of 3 16–bit word types.
TIMe specifies to trigger on either the RT (remote terminal response time), or
the IMG (Inter-message Gap). Use the commands TRIGger:A:BUS:B<x>:
MIL1553B:TIMe:QUALifier, TRIGger:A:BUS:B<x>:MIL1553B:TIMe:
LESSLimit,andTRIGger:A:BUS:B<x>:MIL1553B:TIMe:MORELimit to
specify the time parameters.
ERRor specifies to trigger upon a signaling error. (You can specify which type
of error Parity, Sync, Manchester or Non-contiguous Data by using the
TRIGger:A:BUS:B<x>:MIL1553B:ERRTYPE command.)
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-837