Commands Listed in Alphabetical Order
Examples
WFMInpre:WFMT
YPe RF_FD
specifies the type of the incoming waveform as a
frequency domain trace (to be displayed on the frequency domain window).
WFMInpre:WFM
TYPe?
might return ANALOG, indicating that the incoming
waveform type is either Channel 1–4 or Math (to be displayed on the time domain
window).
WFMInpre:X
INcr
This command specifies the horizontal interval between incoming waveform
points sent to the oscilloscope using the
CURVE command. The units are time,
in second
s, or frequency, in hertz, and can be specified or queried using the
WFMInpre:XUNit command.
(See pag
e 2-95, Waveform Transfer Command Group.)
Group
Waveform Transfer
Syntax
WFMInpre:XINcr <NR3>
WFMInpre:XINcr?
Related Commands
WFMInpre:XUNit,
WFMOutpre:XINcr?
Arguments
<NR3> is a floating point number that specifies the horizontal interval
representation.
Examples
WFMINPRE:XINCR 3E-3 sets the interval between incoming waveform points
to3ms.
WFMINPRE:XINCR ? might return :WFMIN PRE:XINCR 1.0000E-3 indicating
that if WFMInpre:XUNit is set to "s", there is a 1 ms interval between incoming
waveform points.
WFMInpre:XUNit
This command specifies the horizontal units of the x-axis of the data points being
sent to the oscilloscope using the
CURVE command. This value can be in “s”
for seconds, or “Hz”.
(See page 2-95, Waveform Transfer Command Group.)
2-954 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual