Performance Verication
Check Residual Spurious
Response
This check veri
es that the oscilloscope m eets the specication for residual
spurious response. This check does not require an input signal.
1. Terminate the
oscilloscope RF input in 50 and make sure that no input
signal is applied.
2. Set the reso
lution bandwidth (RBW) as follows: Push the front-panel BW
button. Push the side-bezel RBW Mode button to select Manual. Push the
side-bezel RBW button. Set the resolution bandwidth to 50 kHz.
3. Turn auto level on as follows:
a. Push the fr
ont-panel Ampl button.
b. Push the side-bezel Auto Level button to turn the auto level feature on.
The instr
ument will automatically set the reference level correctly.
4. Check in the range of 50 kHz to 3 GHz as follows:
a. Push the front-panel Freq/Span button. Push the side-bezel Start button.
Set the start frequency to 50 kHz. Push the side-bezel Stop button. Set the
stop fr
equency to 3 GHz.
b. Observe any spurs that are greater than - 80 dBm and note them in the
test r
ecord.
5. Check in the range of 2.75 GHz to 4.5 GHz as follows:
a. Change the oscilloscope start frequency to 2.75 GHz and the stop
frequency to 4.5 GHz.
b. Observe any spurs that are greater than - 78 dBm and note them in the
test record.
6. Check in the range of 3.5 GHz to 6.0 GHz as follows:
a. Change the oscilloscope start frequency to 3.5 GHz and the stop frequency
to 6.0 GHz.
b. Observe any spurs that are greater than - 78 dBm and note them in the
test record.
104 MDO4000 Series Specications and Performance Verication