EasyManua.ls Logo

Tektronix MDO4104 - Check Analog Bandwidth

Tektronix MDO4104
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Ve ri cation
Check Analog Bandwidth
This test check
s the bandwidth at 50 and 1 M for each channel.
1. Connect the output of the leveled sine wave generator (for example, Fluke
9500) to the os
cilloscope channel 1 input as shown in the following
illustration.
WARNING. The generator is capable of providing dangerous voltages. Be sure to
set the generator to off or 0 volts before connecting, disconnecting, and/or moving
the test hookup during the performance of this procedure.
2. Push the front-panel Default Setup button.
3. Select 50 impedance as follows:
a. Set the calibrator to 50 output impedance and to generate a sine wave.
b. Push the front-panel channel 1 button.
c. Set t
he Termination (input impedance) to 50 .
4. Set the Acquisition mode to Sample as follows:
a. Push the front-panel Acquire button.
b. Push the Mode lower-bezel button (if not already selected).
c. Push the Sample side bezel button.
5. Set the Vertical Scale to 1mVper d ivision.
6
.
F
or vertical scales less than 500 mV/div, adjust the signal source to at least
8 vertical divisions at the selected vertical scale with a set frequency of
50 kHz. For example, at 5 mV/div, use a 40 mV
p-p
signal, at 2 mV/div, use a
16 mV
p-p
signal, and at 1 mV/div, use a 8mV
p-p
signal. For vertical scales
of 500 mV/div and 1 V/div adjust the signal source to 3 V
p-p
.Useasinewave
for the signal source.
7. Set the Horizontal Scale to 10 μs per division.
MDO4000 Series Specications and Performance Verication 73

Table of Contents

Related product manuals