Specications
Data Handling
Specications
The following table shows the data handling specications for the MDO4000
Series oscilloscopes.
Table 10: Da
ta handling specications
Characteristic Description
Nonvolatil
e memory retention time,
typical
No time limit for front-panel settings, saved waveforms, setups, or calibration constants.
10 M and 20 M records saved as Reference waveforms are not saved in the nonvolatile
memory and
they will not be saved across a power cycle.
Real-time clock A programmable clock providing time in years, months, days, hours, minutes, and seconds
Power Supply System Specications
The foll
owing table shows the power supply syste m specications for the
MDO4000 Series oscilloscopes.
Table 11: Power supply system specications
Characteristic Description
Operating line frequency and voltage
range
Volts: 100 240; Hz: 50 60
Volts: 115: Hz: 400
Maximum power consumption: 225 W
Source voltage 100 V to 240 V ±10%
Source frequency (85 to 264 V ) 45 Hz to 66 Hz
(100 V to 132 V) 360 Hz to 440 Hz
Fuse rating
T6.3AH, 250 VAC
The
fuse cannot be replaced by the user.