EasyManua.ls Logo

Tektronix MDO4104 - Check Phase Noise

Tektronix MDO4104
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Ve ri cation
Check Ph ase Noise
This step check
s the phase noise measured offset from a 2 GHz CW signal. It
checks at three offset frequencies: 10 kHz, 100 kHz, and 1 MHz.
WARNING. The generator is capable of providing dangerous voltages. Be sure to
set the generator to off or 0 volts before connecting, disconnecting, and/or moving
the test hookup during the performance of this procedure.
NOTE. Do not use an N connector with the Anritsu generator. Equipment damage
will result if an N connector is used.
1. Connect the output of a signal generator, such as the Anritsu generator, to the
oscilloscope RF Input, using a 50 SMA coaxial cable (see the following
gure).
2. Set the generator for a 2 GHz, 0 dBm signal.
3. Initial oscilloscope setup:
a. Push the front-panel Default Setup button.
b. Turn Channel 1 off.
c. Push the front-panel RF button to turn on the RF channel and display
the bottom-bezel RF menu.
d. Turn on the average trace as follows: Push the bottom-bezel Spectrum
Traces button. Push the side-bezel Average button to set the Average
Traces to On. Push the side-bezel Normal button to set Normal to Off.
e. Set the center frequency to 2 GHz as follows: Push the front-panel
Freq/Span button. Push the side-bezel Center Frequency button. Set the
center frequency to 2 GHz.
f. Set the span to 50.0 kHz as follows: Set the Span to 50.0 kHz.
g. Center the signal on the display. To do this:
MDO4000 Series Specications and Performance Verication 85

Table of Contents

Related product manuals