Performance Verication
Random Noise, Sam ple Acquisition Mode
Vertical sensitivity = 100 mV/div
Performance checks 1 M 50
Bandwidth Test result (mV) High lim it (mV) Test result (mV) High limit (m V)
MDO4104-3, MDO4104-6
Full 8.30 6.08
250 MHz limit 5.10 4.05
Channel 1
20 MHz limit 5.10 4.05
Full 8.30 6.08
250 MHz limit 5.10 4.05
Channel 2
20 MHz limit 5.10 4.05
Full 8.30 6.08
250 MHz limit 5.10 4.05
Channel 3
20 MHz limit 5.10 4.05
Full 8.30 6.08
250 MHz limit 5.10 4.05
Channel 4
20 MHz limit 5.10 4.05
MDO4054-3, MDO4054-6
Full 8.13 8.13
250 MHz limit 6.10 6.10
Channel 1
20 MHz limit 4.10 4.10
Full 8.13 8.13
250 MHz limit 6.10 6.10
Channel 2
20 MHz limit 4.10 4.10
Ful
l
8.1
3
8.1
3
250 MHz limit 6.10 6.10
Channel 3
20 MHz limit 4.10 4.10
Fu
ll
8.
13
8.
13
250 MHz limit 6.10 6.10
Channel 4
20 MHz limit 4.10 4.10
44 MDO4000 Series Specications and Performance Verication