Performance Verication
DeltaTimeMeas
urement Accuracy
MDO4104-3, MD
O4104-6
Channel 2
MDO = 4 ns/div
, Source freq = 240 MHz
MDO V/div S ou rce V
pp
Test r esult High limit
5mV 40mV 118ps
100 mV 800 mV 117 ps
500 mV 4 V 117 ps
1 V 4 V 122 ps
MDO = 40 ns /div, S ou rce freq = 24 MHz
1 mV 8 mV 464 ps
5 mV 40 mV 276 ps
100 mV 800 mV 234 ps
500 mV 4 V 232 ps
1 V 4 V 417 ps
MDO = 4
00 ns/div, Source freq = 2.4 MHz
1 mV 8 mV 4.50 ns
5mV 40mV 2.52ns
100 m
V
800 m
V
2.05
ns
500 mV 4 V 2.03 ns
1 V 4 V 4.01 ns
MDO = 4 μs/div, Source freq = 240 kHz
1 mV 8 mV 45.
0ns
5mV 40mV 25.2ns
100 mV 800 mV 20.5 ns
50
0mV
4V 20
.3 ns
1V 4V 40.1ns
MDO = 40 μs/div, Source freq = 24 kHz
1 mV 8 mV 450 ns
5 mV 40 mV 252 ns
100 mV 800 mV 205 ns
500 mV 4 V 203 ns
1 V 4 V 401 ns
MDO = 400 μs/div, Source freq = 2.4 kHz
1 mV 8 mV 4.50 s
5mV 40mV 2.52s
100 mV 800 mV 2.05 s
500 mV 4 V 2.03 s
1 V 4 V 4.01 s