Performance Verication
DeltaTimeMeas
urement Accuracy
MDO4054-3, MD
O4054-6, MDO4034-3, MDO4014-3
Channel 4
MDO = 4 ns/div
, Source freq = 240 MHz (except for the MDO4014-3)
MDO V/div S ou rce V
pp
Test r esult High limit
5 mV 40 mV 234 ps
100 mV 800 mV 233 ps
500 mV 4 V 233 ps
1 V 4 V 237 ps
MDO = 40 ns /div, S ou rce freq = 24 MHz
1 mV 8 mV 736 ps
5 mV 40 mV 423 ps
100 mV 800 mV 357 ps
500 mV 4 V 354 ps
1 V 4 V 581 ps
MDO = 4
00 ns/div, Source freq = 2.4 MHz
1 mV 8 mV 6.99 ns
5mV 40mV 3.54ns
100 m
V
800 m
V
2.73
ns
500 mV 4 V 2.69 ns
1 V 4 V 5.34 ns
MDO = 4 μs/div, Source freq = 240 kHz
1 mV 8 mV 69.
9ns
5mV 40mV 35.4ns
100 mV 800 mV 27.3 ns
50
0mV
4V 26
.9 ns
1V 4V 53.4ns
MDO = 40 μs/div, Source freq = 24 kHz
1 mV 8 mV 699 ns
5 mV 40 mV 354 ns
100 mV 800 mV 273 ns
500 mV 4 V 269 ns
1 V 4 V 534 ns
MDO = 400 μs/div, Source freq = 2.4 kHz
1 mV 8 mV 6.99 s
5mV 40mV 3.54s
100 mV 800 mV 2.73 s
500 mV 4 V 2.69 s
1 V 4 V 5.34 s