Performance Verification
8. Repeat the test
sat50Ω as follows:
a. Set the calibrator impedance to 50 Ω.
b. Set the Termination (input impedance) to 50 Ω.
c. Repeat steps 4 through 6.
9. Repeat the procedure for all remaining channels as follows:
a. Push the front-panel channel button to deselect the channel that you
already tested.
b. Connect the calibrator to the input for the next channel to be tested.
c. Starting
from step 3, repeat the procedure for each channel.
64 MDO4000 Series Specifications and Performance Verification