SPI serial bus trigger settings panel
NOTE. Requires option SREMBD.
Field or control Description
Source Select the SPI bus on which you want to trigger.
Trigger On Select the type of information on which to trigger.
Data Words Sets the number of data words (1 word = 8 bits) on which to trigger (one to sixteen bytes). Use
the A knob to change the value.
Available when Trigger On = Data.
Data Sets the data pattern on which to trigger. The number of bits shown depends on the Data
Words setting.
Tap the Binary or Hex field and use the A and B knobs to select and change the values. Or
double-tap on the field and use the virtual keypad to enter values.
Available when Trigger On = Data.
A, B knob controls Use the A knob to select (highlight) the digit(s) to change.
Use the B knob to change the value of the digit(s).