TCPA300 and TCPA400 Performance Verification
TCPA400 Test R
ecord
Photocopy this form and use it to record the performance test results.
TCPA400 test record
Instrument S erial Number: Certificate Number:
Temperature
:
Relative Humidity %:
Date of Calibration:
Technician
:
Performance test Minimum Measured/calculated Maximum
Front-Pan
el Display
Pass/Fail
DC Gain Accuracy (%
Error)
-1% +1%
Bandwidth .707
——
AC Coupling
1
TEKPROBE Cable
1050 Ap-
p
1500 Ap-
p
BNC Coax Cable
1.05 Vp-
p
1.5 Vp-p
Degaus
s
Pass/Fail
1
The AC Coupling test can be performed with either cable, but the units of measure for the results differ between the two.
18 TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Service Manual